Before You Use GoQuick
• Get training from your doctor or nurse.
• Know your dose. Know the pen parts.
• Make sure yo have the pen with the blue injection button.
• Wash your hands.
Setting Up and Using a New GoQuick
Step 1. Attach the Needle
a. Pull the white pen cap straight off the pen.
b. Peel the seal from a new needle.
c. Firmly grasp the cartridge holder.
d. Push the needle onto the cartridge holder tip.
e. Gently screw the needle onto the pen. Do not overtighten.
f. Leave both needle covers on the needle.
Step 2. Mix the Genotropin
a. Hold the pen with the needle-end pointing up and the A facing you.
b. Firmly twist the cartridge holder into the pen until B clicks into the notch.
• Gently tilt the pen from side to side. Do not shake the pen. Shaking
may damage the growth hormone.
c. Check that the liquid in the cartridge is clear. All the powder should be dissolved.
• If not, gently tilt the pen from side to side a few more times.
d. Check the liquid again. Make sure it is clear.
• If the liquid is clear, go to Step 3.
• If the liquid is still cloudy or you see any powder, use a new pen.
Step 3. Remove the Air
a. Pull the outer needle cover off. Save it to remove the needle.
b. Leave the inner needle cover on.
c. Hold the pen with the needle-end pointing up.
d. Tap the cartridge holder gently to help any trapped air move to the top.
e. Firmly, twist the cartridge holder into the pen until C clicks into the notch.
• Some liquid may appear around the inner needle cover.
Step 4. Attach the Needle Guard (Optional)
a. Pull the black cap off the needle guard.
• If the needle shield slides out, push it back into the needle guard until it clicks into place.
b. Hold the pen in one hand below the blue logo. With the other hand, hold the needle
guard below the needle shield.
c. Line up the black logo on the needle guard with the blue logo on the pen. Carefully push
the needle guard onto the pen until it snaps into place.
Step 5. Prime the Pen
a. Pull the inner needle cover off. Throw it away.
b. For 5.3 mg Pen, check that 0.1 mg is set in the memory window. For 12 mg Pen, check that 0.3 mg is set in the memory window.
c. Turn the grey dial in the direction of the arrows until it stops clicking.
d. Hold the pen with the needle pointing up.
e. Push the blue injection button until liquid appears.
f. If liquid does not appear at Step “e”, repeat Steps b-e in this section up to two more times.
g. If liquid still does not appear, do not use the pen.
• See the Questions and Answers section below for more information.
h. If you use the needle guard, press the black button to release the needle shield.
Step 6. Set the Dose
• Use the black ring to set the dose. Be careful not to turn the grey dial while setting the dose.
a. Hold the black ring as shown in.
b. Turn the black ring until your dose lines up with the white pointer. Your doctor or nurse
has told you your dose.
c. If you turn your dose past the white pointer, just turn the black ring back to set the correct dose.
d. Once you have set your dose, do not change it unless your doctor or nurse tells you.
Step 7. Draw Up a Dose
a. Turn the grey dial in the direction of the arrow until the clicking stops.
b. Your dose on the black rod should line up with the white pointer.
c. Check that the dose you drew up on the black rod is the same as the dose you set in
the memory window. shows an example.
d. If the doses do not match, make sure you have turned the grey dial in the direction of
the arrow until it does not click anymore.
Step 8. Give the Injection
a. Prepare an injection site as your doctor or nurse has told you.
b. Hold the pen over the injection site.
c. Push the pen down to insert the needle into the skin.
d. Using your thumb, push the blue injection button down until it stops clicking.
• Count for 5 seconds before you pull the needle out of the skin. Keep light pressure on
the button with your thumb while you count.
e. Pull the pen straight out from the skin.
Step 9. Remove the Needle; Cap and Store Your Pen
Step 9a: With needle guard
a. Place the outer needle cover into the end of the needle shield.
b. Use the needle cover to push in the needle shield until it locks into place.
c. Use the needle cover to unscrew the needle and put it in a proper container for used needles.
d. Leave the needle guard on the pen.
e. Place the black cap on the needle guard. Store your pen in the refrigerator.
Step 9b: Without needle guard
a. Do not touch the needle.
b. Carefully cover the needle with the outer needle cover.
c. Use the needle cover to unscrew the needle and put it in a proper container for used needles.
d. Place the white cap on the pen. Store your pen in the refrigerator.
GH, growth hormone.
Reference:
PP-GEN-IND-0942 June 2023
Simplicity and flexibility in administration1
Proven in different health conditions2-5
Multiple support materials
Review the upcoming events
To access further materials, resources and receive communication about medicines and vaccines promoted by Pfizer.
Lorem ipsum dolor sit amet, consectetur adipiscing elit, sed do eiusmod tempor incididunt ut labore et dolore magna aliqua.
Lorem ipsum dolor sit amet, consectetur adipiscing elit, sed do eiusmod tempor incididunt ut labore et dolore magna aliqua.
Lorem ipsum dolor sit amet, consectetur adipiscing elit, sed do eiusmod tempor incididunt ut labore et dolore magna aliqua. Hac habitasse platea dictumst quisque sagittis purus sit amet volutpat. Lectus magna fringilla urna porttitor rhoncus. Venenatis urna cursus eget nunc scelerisque viverra. Id donec ultrices tincidunt arcu non sodales.
Sagittis aliquam malesuada bibendum arcu vitae elementum curabitur. Pellentesque elit ullamcorper dignissim cras tincidunt. Orci ac auctor augue mauris augue neque. Dui vivamus arcu felis bibendum ut tristique et egestas quis. Sed vulputate mi sit amet mauris commodo.
Nunc eget lorem dolor sed viverra ipsum. Sed ullamcorper morbi tincidunt ornare.
These pages are not intended for patients or for members of the general public. The web pages contain promotional content.
If you select 'No', you will be redirected to Pfizer.co.in
For more details on, Who is a Registered medical practitioner*, please visit "https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf-documents/acts_rules/2016DrugsandCosmeticsAct1940Rules1945.pdf", Page No. 39, Rule 2 part (ee), last accessed on 26th April 2021.
Address: The Capital, A Wing, 1802, 18th Floor, Plot No. C-70, 'G' Block, Bandra Kurla Complex, Bandra East, Mumbai - 400051.